GreenLight XPS™

Laser Therapy System

Console: XPS

  • Laser type: solid state, frequency doubled
  • Wavelength: 532 nm
  • Max. power output: 180 W
  • Nominal optical hazard distance (NOHD): 33.9 meters (MPE = 1x10-3 W/cm2)
  • Eye protection: OD ≥ 6
  • Working distance: 0.5−3.0 mm
  • Vaporization beam mode: quasi-CW (15 kHz−25 kHz)
  • Coagulation beam mode: pulse modulated by 12 Hz, 25% duty cycle
  • Compatible fibers: MoXyTM fiber (750 µm core diameter), HPS fibers (600 µm core diameter)
  • Optical penetration: 0.8 mm
  • Coagulation depth: 1−2 mm
  • Coagulation mode: yes; TruCoagTM 5−40 W
  • FiberLifeTM: yes
  • MoJoTM Software enabled: yes
  • Current draw: 20 A
  • Cooling water: internal
  • Dimensions: 20” W x 36” D x 43.5” H (50.8 cm x 91.4 cm x 110.5 cm)
  • Weight: approximately 420 lb (190 kg)
  • Length of power cord: 15 ft (4.6 m)

Fiber: MoXy

  • Laser compatibility: XPS
  • Firing angle: ~70−80°
  • Power range: ≤ 180 W
  • Beam area at working distance: 0.44 mm2
  • Energy limit: 650 kJ
  • Fiber core diameter: 750 µm
  • Fiber tip O.D.: 2.3 mm
  • Active Cooling CapTM: Yes
  • Use environment: most 22−26 FR continuous flow endoscope/cystoscope