AGENT™ Paclitaxel-Coated Balloon Catheter 

Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician. 

INTENDED USE / INDICATIONS FOR USE

The AGENT Paclitaxel-Coated Balloon Catheter is intended to be used after appropriate vessel preparation in adult patients undergoing percutaneous coronary intervention (PCI) in coronary arteries 2.0 mm to 4.0 mm in diameter and lesions up to 26 mm in length for the purpose of improving myocardial perfusion when treating in-stent restenosis (ISR).

CONTRAINDICATIONS

Use of the AGENT Drug-Coated Balloon Catheter is contraindicated in the following: 

  • Use in the supra-aortic/cerebrovascular arteries. 
  • Unprotected native left main coronary artery disease. 
  • Coronary artery spasm in the absence of a significant stenosis. 
  • Patients with known hypersensitivity to paclitaxel (or structurally-related compounds). 
  • Patients who cannot receive recommended antiplatelet and/or anticoagulant therapy. 
  • Pregnant or breast-feeding women or women who are intending to become pregnant, or men intending to father children. 

WARNINGS 

  • The drug-coated balloon catheter should only be used by physicians experienced in percutaneous coronary intervention.
  • As for all PCI procedures, the need for on-site surgical backup should be considered per the 2023 SCAI Expert Consensus Statement on Percutaneous Coronary Interventions Without On-Site Surgical Backup.
  • As for all PCI procedures, use in patients who are not acceptable candidates for open heart surgery requires careful consideration, including possible hemodynamic support during the procedure, as treatment of this patient population carries special risk.
  • Use extreme caution and careful judgment in patients who have severe reaction to contrast agents that cannot be adequately pre-medicated.
  • Use only the recommended balloon inflation medium (50:50 mixture of contrast medium and sterile saline solution). Never use air or any gaseous medium to inflate the balloon.
  • When the balloon catheter is exposed to the vascular system, it should be manipulated using high-quality fluoroscopic observation.
  • Do not advance or withdraw the catheter unless the drug-coated balloon is fully deflated under vacuum. If unusual resistance is felt during manipulation, determine the cause of the resistance before proceeding. If the source of resistance cannot be determined, it is recommended to extract the entire system (as a unit) with the guide catheter.
  • If difficulty is experienced during drug-coated balloon treatment, do not continue. Deflate the device and remove the catheter.
  • To reduce the potential for vessel damage, the inflated diameter and length of the balloon should approximate the diameter and length of the vessel just proximal and distal to the stenosis.
  • Do not exceed the rated burst pressure indicated on the compliance chart. Use of an inflation device is recommended to assure accurate pressurization.
  • If difficulty is experienced removing the device, other retrieval methods (additional wires, snares and/or forceps) may cause vessel trauma. Complications can include but are not limited to bleeding, hematoma, or pseudoaneurysm.
  • The safety and effectiveness of using more than one AGENT DCB to treat a single lesion or multiple lesions/vessels per procedure has not been established. Preclinical studies suggest that use of multiple DCBs within a single coronary vessel may be associated with downstream perivascular inflammatory injury as a consequence of microvascular obstruction.

PRECAUTIONS  

  • Any use for procedures other than those indicated in these instructions is not recommended. 
  • Use the drug-coated balloon catheter prior to the “Use By” date specified on the package. 
  • The treatment site should be adequately prepared (intravascular imaging guidance recommended) to achieve the maximum lumen diameter prior to treatment with AGENT Drug-Coated Balloon.

Note: Current US and European PCI guidelines recommend intravascular imaging (IVUS) to identify the mechanism of stent failure. 

  • Drug-coated balloon PCI should be used with caution during procedures involving calcified vessels due to the abrasive nature of these lesions. 
  • Significant stenosis (> 50 %) proximal to the treatment site must be pretreated to prevent abrasion and delamination of the balloon’s drug coating while crossing the lesion. 
  • Precautions to prevent or reduce thrombosis should be taken during any PCI procedure: 
    • The patient should be treated with heparin or similar agent during the procedure. 
    • Flush all products entering the vascular system with heparinized sterile isotonic saline or a similar solution prior to use. 
  • Carefully inspect the drug-coated balloon catheter prior to use to verify it has not been damaged during shipment or preparation; confirm that its size, shape, and condition are suitable for the procedure. 

Note: After removing the balloon protector, a white powdery substance may be observed inside the balloon protector and will not impact the delivered drug dose.  

  • If unusual resistance is felt during removal of the balloon protector or mandrel, do not use the AGENT device and replace with another. 
  • Do not expose the drug-coated balloon catheter to organic solvents such as alcohol or detergents. 
  • When loading or exchanging the balloon catheter, it is recommended to thoroughly wipe the guidewire with heparinized saline to improve catheter movement on the guidewire. 
  • Do not touch, wipe, bend, or squeeze the drug-coated balloon, or allow it to come into contact with any liquids prior to insertion as damage to the device's drug coating or premature release of the drug may occur. 
  • If using a Tuohy-Borst type adapter, take care to not over-tighten the hemostatic valve around the catheter shaft as lumen constriction may occur, affecting inflation/deflation or damaging the device's drug coating. 
  • Never advance the drug-coated balloon catheter without the guidewire extending from the tip. Do not use the AGENT DCB if the lesion is unable to be crossed with a guidewire. 
  • The AGENT DCB is indicated for lesions up to 26 mm using one balloon. The safety and effectiveness of treating lesions longer than 26 mm has not been established. When clinically warranted, observe the following precautions to prevent local drug overdosing: 
    • If treating a long lesion (longer than the maximum balloon length available), each individual segment should be treated only once with a drug-coated balloon. Treat each segment with a new balloon and try to minimize overlapping the treated segments. 
    • Do not use a second drug-coated balloon at the same treatment site. 
  • The AGENT DCB was not studied clinically with use of bail out stenting, including drug-eluting stent (DES). Implanting a drug-eluting stent at the same treatment site may have added risks, including overdosage or interaction between the active agents. 
  • Published literature has reported paclitaxel caused cell aneuploidy at tissue concentrations similar to tissue concentrations recorded after treatment with an AGENT Drug-Coated Balloon in preclinical studies. The aneugenic effect is due to paclitaxel’s pharmacodynamic action of interfering with microtubule disassembly, which is also the basis for the pharmacodynamic action preventing restenosis in vascular tissue following treatment with an AGENT Drug-Coated Balloon. The relevance of both this observation and the aneugenic mechanism of genotoxicity for human carcinogenicity is currently not known. 
  • The AGENT Drug-Coated Balloon has not been tested in pregnant women or in men intending to father children; effects on the developing fetus have not been studied. The risks and reproductive effects of paclitaxel administered through the coronary vasculature remain unknown. The AGENT Drug-Coated Balloon is contraindicated for use in women who are pregnant or attempting to conceive, or men intending to father children. It is not known whether paclitaxel is distributed in human milk. In lactating rats given paclitaxel, milk concentrations appeared to be higher than maternal plasma levels and declined in parallel with the maternal levels.  
  • Possible interactions between paclitaxel and concomitantly administered medications have not been formally investigated. Drug interactions of systemic chemotherapeutic levels of paclitaxel with possible concomitant medications are outlined in the labeling for finished pharmaceuticals containing paclitaxel, such as TAXOL. 
  • The optimal duration of dual anti-platelet therapy (DAPT) post PCI for ISR is currently unknown. Published expert consensus opinion has suggested that DAPT should be used for a minimum of one month post ISR PCI. While the AGENT DCB IDE required that enrolled subjects receive a minimum of 1-month of DAPT followed by antiplatelet monotherapy for the duration of the study, more than 70 % of enrolled patients remained on DAPT for at least 1 year. In conjunction with this information, operators should use clinical judgment in deciding on the duration of DAPT following use of AGENT DCB.   

ADVERSE EVENTS 

Potential adverse events which may be associated with the use of a drug-coated balloon (DCB) or DCB procedure include, but are not limited to, the following:

  • Additional, possibly surgical, intervention
  • Allergy (drug coating and its components, device, medications, contrast)• Arrhythmia including conduction system disorder.
  • Bleeding (including hemorrhage or hematoma possibly requiring transfusion or additional intervention)
  • Cerebrovascular accident (stroke)/transient ischemic attack (TIA)
  • Death
  • Embolism (tissue, plaque, thrombus, device, drug coating)
  • Fever/inflammation• Hemodynamic instability
  • Hypotension/hypertension (shock)
  • Kidney injury/failure
  • Myocardial ischemia/infarction
  • Organ insufficiency/failure (heart, liver, lungs)
  • Pain (anginal, non-anginal)
  • Pericardial effusion/cardiac tamponade
  • Radiation injury
  • Sepsis/infection
  • Slow flow/no reflow
  • Vessel injury (spasm, dissection, perforation, rupture, arteriovenous fistula, aneurysm)
  • Vessel occlusion (abrupt closure, slow flow/no reflow, thrombosis, restenosis)

Potential adverse events not captured above that have been associated with administration of paclitaxel at systemic doses, include, but are not limited to, the following:

  • Abnormal liver enzymes
  • Allergic/immunologic reaction to drug (paclitaxel or structurally-related compounds)
  • Alopecia
  • Anemia
  • Blood product transfusion
  • Gastrointestinal symptoms
  • Hematologic dyscrasia (including leukopenia, neutropenia, thrombocytopenia)
  • Hepatic enzyme changes
  • Histologic changes in vessel wall, including inflammation, cellular damage or necrosis
  • Myalgia/arthralgia
  • Peripheral neuropathy
 
Top