Skip to main content

My account

Boston Scientific accounts are for healthcare professionals only.

Create an account to access online training and education on EDUCARE, manage your customer profile, and connect with customer support and service teams.

My Boston Scientific account

Access your online applications and manage your customer profile.

Quick Links

Contact a representative Online education and training Order products online Reimbursement resources Return a product Customer support

Call customer care

Technical Information

DYNAGEN™ X4 CRT-D

Models G156 and G158

Pacing Therapy

  
Brady Modes 

Normal: DDD(R), DDI(R), VDD(R), VVI(R), AAI(R), Off

 

Temporary: DDD, DDI, DOO, VDD, VVI, VOO, AAI, AOO, Off

AT/AF Management ATR Mode Switch, Ventricular Rate Regulation (VRR) - MIN, MED, MAX, Atrial Flutter Response (AFR), PMT Termination, Rate Smoothing
Rate Adaptive PacingAccelerometer with sensor trending function


Heart Failure Therapy

  
Heart Failure Therapy Optimization

Normal: DDD(R), DDI(R), VDD(R), VVI(R), AAI(R), OffSmartDelay™ (AV Optimization), BiV Trigger, LV Offset, LV Sensing

LV Lead OptionsQuadripolar LV lead
LV Pacing Vector Options17 options
LV Sensing Vector Options8 options plus OFF


Patient Diagnostics

  
Arrhythmia Logbook

Events summary, Stored Electrograms with Annotated Markers, (Intervals and approximately 17 minutes of multichannel EGM, always with 10 seconds Onset and event storage prioritization). Implant activation of all available EGMs. On screen measurement of all stored signal amplitudes and timing

Histograms & CountersTachy Events and Brady Counters
Heart Rate Variability(HRV)SDANN and HRV Footprint (24 hour heart rate collection period)
Daily Trends For Last 365 Days

Events, Activity Level, Atrial Burden, Respiratory Rate, Heart Rate, SDANN, HRV Footprint, Autonomic Balance Monitor (ABM), Lead impedances and amplitudes

 

To note: Automatic activation of all available daily trends at implant.

AT/AF Diagnostics % Atrial Burden, Daily burden, Average V-rate during ATR Mode Switch Episode
Heart Failure Therapy/Diagnostics

HF Perspectiv™ report, Respiratory Rate Trend, Weight, Blood Pressure, Heart Rates, HRV Footprint, SDANN, Atrial Arrhythmia Burden, Activity Level, A & V Arrhythmias, Pacing Histograms

 

To note: Weight and Blood Pressure are only available via LATITUDE.


Device Testing/Induction Methods

  
Induction Methods 

Vfib Induction, Shock on T Induction, Programmed Electrical Stimulation (PES), 50 Hz/Manual Burst Pacing

Commanded Therapy MethodsCommanded Shock, Commanded ATP


Tachyarrhythmia Therapy

  
Sensing/Detection

Zones VF only, or VF and VT or VF, VT, VT-1. Lowest Zone can be Monitor Only

Shock Reduction and Appropriate TherapyAcuShock™ Advanced Technology including Onset/Stability™, RhythmID™ with RhythmMatch™, Dynamic Noise Algorithm (DNA) for sensing, Automatic Gain Control (AGC) with programmable sensing floor, Narrow Band Pass Filter
Antitachycardia Pacing Therapy (ATP) TerminationQuick Convert™ in VF Zone. Two programmable ATP schemes in both VT and VT-1 zones. Burst, Ramp, Scan, Ramp-Scan
Shock Energy 41 J stored, 35 J delivered. First two shocks in each zone programmable. VT-1 has 5 shocks, VT has 6 shocks and VF has 8 shocks. Reverse Last Shock Polarity in zone. Programmable RV Coil to RA Coil and Can (TRIAD), RV Coil to Can, RV Coil to RA Coil (COLD CAN)
Nominals

VF Zone (200 bpm)–Detection: Rate and Duration, Therapy: Quick Convert, 8 high energy shocks

 

VT Zone (160 bpm)–Detection: RhythmID or Onset/Stability, Therapy: ATP x 2, 6 high energy shocks


Implant/In-Clinic Follow Up

  
Implant Communication Mode

Programmable values: Enable use of ZIP™ telemetry (MICS) (Requires initial use of wand for device ID) or use wand for all telemetry

 

Nominal: Enable use of ZIP telemetry (Requires initial use of wand for device ID)

In Clinic Follow-UpWireless ECG


Remote Follow Up

  
Patient Triggered Monitor (PTM)Triggers the storage of two minutes onset and one minute post - EGMs, intervals, and annotated marker data during a symptomatic episode by placing a magnet over the device
Beeper Feature(Patient Alerts)Beep During Capacitor Charge, Beep when Explant is Indicated, Beep when Lead Impedance measurement(Shock or Pace) is Out-of-Range
Magnet FeatureMagnet Response (Off, Store EGM, Inhibit Therapy)
Remote MonitoringThis device is designed to be LATITUDE™ enabled; LATITUDE availability varies by region
WirelessRemote follow-up for all devices (MICS)
Top